Victoreen CVD 700-6B Power Supply Fix, Buzzing Noise.

Similar to my other video where the meter fixed itself, I found the system to have a louder then normal hum and the meter would just go full scale when turned on.Based on earlier research I assumed part of the high voltage power supply had failed. I ordered a "kit" to upgrade the power supply.

Basically just replace the high voltage switching diode and the vacuum tube regulator (Corotron).These replacements were successful and the meter now works again.

See my other Victoreen 700-6B video for more in depth images etc of the system.